Kromasil CelluCoat

Stable performance

Switching between compatible organic mobile phases will not lead to any degradation in performance

Performance after mobile phase changes

“Make every day's work easier”

You can run Kromasil CelluCoat with maintained stability and performance when switching between compatible normal or polar organic phases. Figure 1 illustrates the impact of a mobile phase switch from heptane/2-propanol (90/10) to 100% acetonitrile and back, then to 100% 2-propanol and back. Performance is measured with trans-stilbene oxide as the efficiency on the first eluted peak (Knox plot) and the selectivity. Neither seems hardly affected by the mobile phase switches.

cellucoat_stability_fig01

Fig. 1 - Stability of Kromasil CelluCoat. Efficiency (Knox plot) and selectivity.
Conditions:
Solute: trans-Stilbene oxide
Stationary phase: Kromasil CelluCoat, 5 μm
Mobile phase: heptane/2-propanol (90/10)
Column size: 4.6 x 250 mm
Flow rates: 0.1-1.2 ml/min
Temperature: 25 °C

· Eka Chemicals, Separation Products · SE–445 80 Bohus, Sweden · Phone +46 31 58 70 00 · Fax +46 31 58 77 27 ·