Chiral technical evidence

 
 

Loadability of Kromasil CelluCoat

“Works all
the way”

Kromasil CelluCoat is produced in particle sizes of 3, 5 and 10 µm. Since the manufacturing technology is identical for all particle sizes, any development study made on the smaller particles can be translated to a larger column with 10 µm particles.

Examples:

Binaphthol
Analytical injection of Binaphtol on CelluCoat Circle right-arrow Preparative injection of Binaphtol on CelluCoat   
Analytical injection
Eluent: heptane/2-propanol (90/10)
Sample size: 10 µg
Detection: UV 223 nm
Preparative injection
Eluent: heptane/2-propanol (90/10)  
Sample size: 1 mg
Detection: UV 350 nm
 
  Enantiomer 1 Enantiomer 2
Purity [%] 95.4 97.8
Yield [%] 91.7 72.8
Common conditions:
Phase: Kromasil CelluCoat, 10 µm
Flow rate: 1 ml/min
Column size: 4.6×250 mm
Temperature: 25 °C
Equivalent loading on a
50×250 mm column:
120 mg
Metoprolol
Analytical injection of metoprolol on CelluCoat Circle right-arrow Preparative injection of metoprolol on CelluCoat   
Analytical injection
Eluent:
heptane/2-propanol/DEA (90/10/0.1)
Sample size: 10 µg
Detection: UV 223 nm
Preparative injection
Eluent:
heptane/2-propanol/DEA (90/10/0.1)
Sample size: 20 mg
Detection: UV 290 nm
 
  Enantiomer 1 Enantiomer 2
Purity [%] 100 99.7
Yield [%] 97.9 88.1
Common conditions:
Phase: Kromasil CelluCoat, 10 µm
Flow rate: 1 ml/min
Column size: 4.6×250 mm
Temperature: 25 °C
Equivalent loading on a
50×250 mm column:
2.4 g
· Eka Chemicals, Separation Products · SE–445 80 Bohus, Sweden · Phone +46 31 58 70 00 · Fax +46 31 58 77 27 ·