“Works all
the way”
Kromasil CelluCoat is produced in particle sizes of 3, 5 and 10 µm. Since the manufacturing technology is identical for all particle sizes, any development study made on the smaller particles can be translated to a larger column with 10 µm particles.
|
|
|
||||||||||
|
Analytical injection
Eluent: heptane/2-propanol (90/10) Sample size: 10 µg Detection: UV 223 nm |
Preparative injection
Eluent: heptane/2-propanol (90/10) Sample size: 1 mg Detection: UV 350 nm |
|||||||||||
|
||||||||||||
|
Common conditions:
Phase: Kromasil CelluCoat, 10 µm Flow rate: 1 ml/min Column size: 4.6×250 mm Temperature: 25 °C |
Equivalent loading on a
50×250 mm column: 120 mg |
|||||||||||
|
|
|
||||||||||
|
Analytical injection
Eluent: heptane/2-propanol/DEA (90/10/0.1) Sample size: 10 µg Detection: UV 223 nm |
Preparative injection
Eluent: heptane/2-propanol/DEA (90/10/0.1) Sample size: 20 mg Detection: UV 290 nm |
|||||||||||
|
||||||||||||
|
Common conditions:
Phase: Kromasil CelluCoat, 10 µm Flow rate: 1 ml/min Column size: 4.6×250 mm Temperature: 25 °C |
Equivalent loading on a
50×250 mm column: 2.4 g |
|||||||||||