What is the difference between dead volume and dwell volume?

The dead volume can also be called extra column volume and consists of the volume of a HPLC system between the point of injection to the point of detection, but excluding the column itself. Thus it includes the injection volume, the volume of the injector, the volume of the connecting tubing before and after the column, the volume of the end-fittings and frits, and the volume of the detector flow cell. The dead volume can be measured by replacing the column with a zero-volume connector. By injecting a very small sample amount, the time can be measured between the moment of injection and the maximum peak height. This time multiplied by the flow rate gives you a good estimate for the system dead volume.

The dwell volume or gradient dwell volume is responsible for the time delay of the gradient. It describes the volume of a gradient HPLC system between the mixing chamber and the column inlet. This volume does of course also exist under isocratic elution, but in that case, it has no impact on the separation. The gradient dwell volume comprises of the volume of the gradient mixer, the connecting tubing to the pump, the pump head and check-valves, the tubing between the pump and the injector, the injector itself, and the tubing between the injector and the column inlet. When initiating a gradient, the column is not subjected to the change in eluent composition until the gradient has passed the dwell volume. During that time, the column is operated under isocratic elution. Attention has to be given when transferring a gradient method from one instrument to another. If the dwell volume differs, the retention times will likely differ as well, despite an identical method and column. The dwell volume can be measured by running a step gradient from 100% methanol to 100% methanol + 10 mg/L acetophenon. The UV detector will detect an S-shaped detector trace. The dwell volume is equal to the time between the injection and half height of the detector trace, multiplied by the flow rate.