Javascript message

This site uses javascript. If javascript is disabled in your browser, some functionality may be impaired.

Acetaminophen (paracetamol)

Analysis of acetaminophen (paracetamol) on Kromasil C18 according to USP39-NF34

  • 1 = acetaminophen
    s00687
chromatogram for g1077
Column:Kromasil 100-3.5-C8 4.6 × 100 mm
Part number:MH3CMA10
Eluent A:acetonitrile / potassium phosphate buffer in water (1.7 g/l KH2PO4 + 1.8 g/l K2HPO4, pH 3.2) / triethylamine (50/50/0.2)
Eluent B:methanol
Gradient:0 min: 1%, 3 min: 1%, 7 min: 81%, 7.1 min: 1%, 10 min: 1% methanol
Flow rate:1 ml/min
Temperature:ambient
Detection:UV @ 254 nm

Reference

Application by Beijing Zhenxiang Industry & Commercial Company

Kromasil columns available with the stationary phase used in this application

Click/touch the dot for the combination of stationary phase and column diameter of your interest to access column length availability and further options.
Column diameters, [mm]
FamilyPhasedp [μm]2.13.04.04.610
100 ÅC83.5 MH3CMD__ MH3CMC__ MH3CMB__ MH3CMA__ MH3CMP__